Home / product / Vacuum Components / Vacuum Pump / Turbo Molecular Pump UTM-MI Series
Model | UTM1200A | UTM1600A | UTM2300A | UTM3400A | ||
Inlet port | VG150 ISO160F | VG200/VG250 ISO200F/ISO250F | VG250 ISO250F | VG300/VG350 ISO320F | ||
Outlet port | KF40 | |||||
Ultimate pressure(*1) | Pa | 10-7 | ||||
Pumping speed (*2)(*3) | N2 | L/s | 1030 | 1400 | 2100 | 3200 |
Ar | L/s | 980 | 1330 | 2000 | 3100 | |
H2 | L/s | 660 | 750 | 1380 | 2100 | |
Compression rate | N2 | 2×108 | ≧ 1×108 | ≧ 1×109 | ||
He | 7×103 | 5×103 | 1×105 | |||
H2 | 8×102 | 9×102 | 6×103 | |||
Max inlet pressure(*4) | Pa | 26 | 4.5 | 7 | ||
Max outlet pressure(*4) | Pa | 266 | 200 | 270 | ||
Rotation speed | min-1 | 37200 | 33700 | 27600 | ||
Speed variation | Speed is variable between 25% and 100% of the rated speed.( set as 0.1%) | |||||
Start-up time | ≦ 8minutes | ≦ 9minutes | ≦ 16minutes | |||
Mounting position | In any desired direction | |||||
Surface treatment(*5) | None | |||||
Commun ication | Contact | D-sub 25 pin female Contact input and output signal | D-sub 37 pin female Contact input and output signal | |||
Serial | Serial connecter RS232C/RS485 | D-sub 9 pin male RS232C/RS485 | ||||
Controller | Composition | Pump with integrated controller | ||||
Input electric power | Single phase AC200 〜240V ±10%( 50/60Hz ±2Hz) | |||||
Max power | 0.75kVA | 1.2kVA | ||||
Cooling method | Water cooled | |||||
Cooling water(*6) | Flow rate | L/min | 3~4 | |||
Pressure | MPaG | 0.2 ~ 0.4 | ||||
Noise | dB(A) | 57 | 60 | |||
Weight | kg | 43 | 41 | 56 | 94 | |
Recommended backing pump(*7) | ≧ 600L/min | ≧ 1500L/min | ||||
Applicable standard | CE、TUVus | |||||
Standard accessories | O-ring(Only VG)、Dust cap(Outlet port)、Remote-control Connector、Instruction manual(CD-ROM) | |||||
Select parts | AC Cable | AC Conector / AC Cable(5m/10m/15m/20m) | ||||
Special bolt for fixing pump | M10(Non-RoHS / RoHS)/ M12(Non-RoHS / RoHS) | |||||
Special bolt for fixing pump | Need(M10)(*8) | Not Need(*8) | ||||
Option | Gas purge adaptor (KF10) |
*1:digit number is expressed.
*2:Without a protective net.
*3:Calculated values:UTM2300A and UTM3400A
*4:Max inlet pressure and Max outlet pressure cannot be satisfied at same time.
*5:Do not flow reactive gas and corrosive gas. Please contact us when using other gases.
*6:Not condensing
*7:When evacuating 30SCCM
*8:Grade of bolts,Fixing method and etc. are subject to various conditions. Please contact us for the detailed information.
29/11/100 Le Duc Tho Street, Ward 7, Go Vap District Ho Chi Minh City, Vietnam
5/5/42 Thanh Binh Street, Hai Duong Province, Vietnam