Model | ST200-A | ST200-R | ST200-E (scheduled for release) | |
Type | Analog output type | Serial communication type | EtherCAT communication type | |
Applicable ionization gauge sensor head | SWT-16(NW16), SWT-25(NW25) | |||
Connectable sensors | Multi ionization gauge ST200: 1 Pirani vacuum gauge sensor unit SWU10-R: 1 (option) Pressure sensro unit SAU: 1 (option) | |||
Repeatability(N2) | ST200 simple mode: ±2% | |||
Sampling time | 60 ms, 5x moving average | |||
Value output | Output voltage: 0 to 10 VDC, LOG output: 0.75 V/1 decade Pressure conversion equation: V=7.25+0.75×(LogP-2) P=10^{(V-7.25)/0.75+2} | |||
Update time | 60msec | |||
Control input signals | FIL ON/OFF、FIL 1/2、DEGAS ON/OFFOperates with open collector input, negative logic | – | – | |
Control output signals | Sensor error, Setpoint 1/2/3, Emission valid Watch of electric power of filament Rating: 24VMAX、50mAMAX Saturation voltage:1V | – | – | |
Communication type | – | RS232C/RS-485 | EtherCAT | |
LED display | POWER: Power, ERROR LED (Blinking pattern and color change depending on each mode) | |||
Emission current | 2mA (1×10-2Pa or lower), 10μA (1×10-2Pa or higher) | |||
Degas Method | Electron bombard type 2mA, 330V | |||
Maximum pressure of sensor | 2×10+5Pa (1.5 x 10+3 Torr, 2 x 10+3 mbar) (Absolute pressure) | |||
Internal volume of sensor | 17cm3(SWT-16), 19cm3(SWT-25) | |||
Operating temperature range | 10 to 50℃ (50 to 122℉) | |||
Operating humidity range | 15 to 80% (not condensing) | |||
Storage temperature | -20 to 65℃ (-4 to 149℉) (At non-energizing, not condensing) | |||
IP rating | IP30 | |||
Power supply voltage | DC20 to 28V (Ripple, Noise 1% or less) Lower than 8W | |||
Input/output connector | D-sub15-pin (M2.6 screw) | |||
Weight | Sensor unit: approx. 280g, sensor head(SWT-16, SWT-25): 80g | |||
Dimension W x D x H | 69mm×63mm×90mm (sensor unit only) | |||
Applicable standard | CE | |||
Option | Multi Ionization Gauge Sensor head | SWT-16(NW16), SWT-25(NW25) Material of gas contacting sections: Filament 1, 2 -Ir/Y2O3 Other: PtC-Mo, SUS304, W, Kovar glass, Kovar/Ni plating | ||
Pirani Vacuum Gauge | SWU10-R *Previous model SPU also available | |||
Pirani Vacuum Gauge Sensor head | SWP-16(NW16), SWP-25(NW25), SWP-R1/8(R1/8), SWP-P18(Φ18 pipe end), SWP-P15(Φ15 pipe end), SWP-CF16(ICF034) Material of gas contacting sections : Filament-Pt, Other-BS/Ni plating, Ni, Solder | |||
Pressure switch | SAU(NW16) Material of gas contacting sections : SUS316L | |||
Unit cable | GUC-200P (for SWU10-R) :0.5m, 1m, 2m GUC-200A (for SAU) :0.5m, 1m, 2m | |||
Display unit | 1CH:ISG1(DC24V) | |||
Display cable | Cable betweeen ST200 and desiplay unit 2m, 5m, 10m, 15m, 20m, 25m, 30m, 35m, 40m | |||
Connector | I/O connector (D-sub15-soket(M2.6mm screw)) | |||
Others | Test certificate, Calibration certificate |
ST200 series have three options; simple mode with a multi ionization gauge, combination mode with a multi ionization gauge and a pirani gauge (SWU10-R), and triple combination mode with a multi ionization gauge, a pirani gauge (SWU10-R) and a pressure switch (SAU)
Mode | ST200-A/ST200-R/ST200-E Simple mode | ST200-A/ST200-R/ST200-E +SWU10-R Combination mode | ST200-A/ST200-R/ST200-E +SWU10-R+SAU Triple combination mode |
Measurable pressure range | 1.0×10-5 to 1.0×10+1Pa | 1.0×10-5 to 1.0×10+5Pa | 1.0×10-5 to 1.0×10+5Pa |
Accuracy | 1.0×10-4 to 3.0×10+0Pa: ±10% | 1.0×10-4 to 1.0×10+4Pa:±10% 1.0×10+4 to 1.0×10+5Pa:±20% | 1.0×10-4 to 1.0×10+4Pa:±10% 1.0×10+4 to 3.0×10+5Pa:±3%F.S.(3.0×10+3Pa・Base temperature: 25℃) |
Connection image | ![]() | ![]() | ![]() |
Measurable pressure range image | ![]() Triple combination can be used when “atmospheric pressure” check required to open chamber. |
For PC | Download from ULVAC’s website “ULVAC SHOWCASE” download center |
For smartphone | Download from Google Play |
PC | Windows 10 64 bit or later |
Smartphone | Android 6 or later, USB TypeC |
29/11/100 Le Duc Tho Street, Ward 7, Go Vap District Ho Chi Minh City, Vietnam
Cookie | Duration | Description |
---|---|---|
cookielawinfo-checkbox-analytics | 11 months | This cookie is set by GDPR Cookie Consent plugin. The cookie is used to store the user consent for the cookies in the category "Analytics". |
cookielawinfo-checkbox-functional | 11 months | The cookie is set by GDPR cookie consent to record the user consent for the cookies in the category "Functional". |
cookielawinfo-checkbox-necessary | 11 months | This cookie is set by GDPR Cookie Consent plugin. The cookies is used to store the user consent for the cookies in the category "Necessary". |
cookielawinfo-checkbox-others | 11 months | This cookie is set by GDPR Cookie Consent plugin. The cookie is used to store the user consent for the cookies in the category "Other. |
cookielawinfo-checkbox-performance | 11 months | This cookie is set by GDPR Cookie Consent plugin. The cookie is used to store the user consent for the cookies in the category "Performance". |
viewed_cookie_policy | 11 months | The cookie is set by the GDPR Cookie Consent plugin and is used to store whether or not user has consented to the use of cookies. It does not store any personal data. |